Application
Temperature, humidity and depression combined tests on electronics components powered with a
tension of less than 6000 volts.
Description of the equipment
The unit including :
- The test space with its conditioning at the rear.
- The machinery at the back with the electrical cabinet.
- The control board in front of the electrical cabinet.
- Supervision with and control thru a touch panel.
Altitude chambers
· Temperature
Range : -65°C to +200°C
Average cooling rate :
10°C/min from +174°C to -37°C with vacuum
Average heating rate :
10°C/min from -37°C to +175°C with vacuum
Stability of temperature ±0,5°C
Specimen dissipation in test: 2 kW at – 30°C.
· Hygrometry
Range of temperature : from +10°C to +95°C
Range of hygrometry : 10 to 98% HR without
dissipation and vacuum (during 4h max).
Homogeneity : ±3% to 10 cm of the walls.
· Vacuum
Range : Pa with 1mbar
On the range of 100°C with -65°C
· Homogeneity of the chamber
Homogeneity of temperature : ±2°C at 10 cm of the walls.
Homogeneity of hygrometry : ±3% at 10 cm of the walls.